Cobalt-selective electrode based on titanium ditelluride intercalated with cobalt / Velikanova TV,Titov AN,Mityashina SG,Vdovina OV // JOURNAL OF ANALYTICAL CHEMISTRY. - 2001. - V. 56, l. 1. - P. 56-59.

ISSN/EISSN:
1061-9348 / нет данных
Type:
Article; Proceedings Paper
Abstract:
A cobalt-selective solid-contact electrode was prepared on the basis of titanium ditelluride. The electrode exhibits the slope of the electrode function -(29 +/- 1) mV/pC and provides the determination of 1 x 10(-5)-1 x 10(-1) M cobalt(II) in the pH range 4.5-6.5. The electrode was used for the potentiometric indication of the titration end point in the determination of cobalt(II) in lanthanum-strontium cobaltite La0.5Sr0.5CoO3. The intercalant concentration range providing optimal ion-selective properties was determined.
Author keywords:
нет данных
DOI:
10.1023/A:1026771529433
Web of Science ID:
ISI:000166747500012
Соавторы в МНС:
Другие поля
Поле Значение
Month JAN
Note Vth All Russian Conference on Electrochemical Methods of Analysis (EMA-99), MOSCOW, RUSSIA, DEC 06-08, 1999
Publisher MAIK NAUKA/INTERPERIODICA
Address C/O KLUWER ACADEMIC-PLENUM PUBLISHERS, 233 SPRING ST, NEW YORK, NY 10013-1578 USA
Language English
Research-Areas Chemistry
Web-of-Science-Categories Chemistry, Analytical
ResearcherID-Numbers Titov, Alexander/K-4328-2013
ORCID-Numbers Titov, Alexander/0000-0003-3636-2604
Number-of-Cited-References 6
Usage-Count-Since-2013 3
Journal-ISO J. Anal. Chem.
Doc-Delivery-Number 398GA