A speckle-interferometric device for studying the cell biological activity / Malygin A. S.,Bebenina N. V.,Vladimirov A. P.,Mikitas K. N.,Baharev A. A. // INSTRUMENTS AND EXPERIMENTAL TECHNIQUES. - 2012. - V. 55, l. 3. - P. 415-418.

ISSN/EISSN:
0020-4412 / нет данных
Type:
Article
Abstract:
A modernized device intended for diagnosing the temperature-dependent activity of cells is described. The activity of cells was determined by finding the correlation coefficient of optical signals in speckle patterns recorded at different moments of time. It is shown that, as the temperature increases from 26 to 36A degrees C, variations in the optical paths in cells increase from 16 to 26 nm. This is probably associated with the physiological activity of cells-trapping of large molecules or their conglomerates. In contrast to the previous variant of the device design, instead of a lipid membrane, a cell monolayer was used, which was cultivated on a plane-parallel glass plate. In order to reduce variations in the optical path lengths in a nutrient solution, a 1.7-mm-thick plane-parallel glass plate was additionally inserted into the cuvette. The modernization of the device also includes heating of the cuvette with hot air flows.
Author keywords:
нет данных
DOI:
10.1134/S0020441212020091
Web of Science ID:
ISI:000304877000017
Соавторы в МНС:
Другие поля
Поле Значение
Month MAY
Publisher MAIK NAUKA/INTERPERIODICA/SPRINGER
Address 233 SPRING ST, NEW YORK, NY 10013-1578 USA
Language English
Research-Areas Engineering; Instruments \& Instrumentation
Web-of-Science-Categories Engineering, Multidisciplinary; Instruments \& Instrumentation
Author-Email somt@bk.ru
ResearcherID-Numbers Vladimirov, Alexander/K-2967-2012
ORCID-Numbers Vladimirov, Alexander/0000-0001-6343-9772
Number-of-Cited-References 11
Usage-Count-Last-180-days 4
Usage-Count-Since-2013 26
Journal-ISO Instrum. Exp. Tech.
Doc-Delivery-Number 953NY